15 - 90% RH non-condensing All dimensions in mm LIFETIME IMPORTANT NOTE:
Expected Operating Life Two years in air
Connection should be made via PCB sockets Long Term Output Drift <5% signal loss per year only. Soldering to the pins will seriously damage
Storage Life 6 months in CTL container
your sensor and invalidate the warranty. * Specifications are valid at 20°C, 50% RH and 1013 mBar, using City Technology recommended circuitry. Performance characteristics outline the performance of sensors supplied within the first 3 months. Output signal can drift below the lower limit over time. Doc. Ref.: microcelcf.indd Issue 8 ECN I 4249 Page 2 of 3 20th February 2017 The Right Sensor Can Sa
Gas
Concentration Used (ppm) MICROceL CF (ppm CO)
Hydrogen Sulfide, H2S
15 < 0.5
Sulfur Dioxide, SO2
5 ± 0.1
Nitric Oxide, NO
35 <6
Nitrogen Dioxide, NO2
20 ±1
Hydrogen, H2
100 < 40
Ethanol, C2H5OH
200 ±1 SAFETY NOTE This sensor is designed to be used in safety critical applications. To ensure that the sensor and/or instrument in which it is used, are operating properly, it is a requirement that the function of the device is confirmed by exposure to target gas (bump check) before each use of the sensor and/or instrument. Failure to carry out such tests may jeopardize the safety of people and